WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

6.5 Zero point and/or offset correction
Zero adjustment for overpressure instruments

If the measured value shown on the CPH6000, with a positive/relative pressure sensor

connected and the test assembly vented to atmosphere, is not equal to zero, then by

pressing the CLEAR key twice (within five seconds), the zero point can be corrected

(maximum allowable correction value is twice the magnitude of the accuracy class).

Offset correction for absolute pressure sensors

For absolute pressure sensors, an offset correction can be made via the menu (see

chapter 7.4.3 "SETUP additional menu items: ReferenceSensor")

7. Commissioning, operation

Instrument features

The instrument features 3 operating modes: MEASURING / CALIBRATION / SWITCH-

TEST, which each offer the user the maximum possible convenience in relation to the

application. To power the test items and to read their measurement signals, there are

various electrical inputs and outputs, which, as an option, can be shielded from harsh

field conditions by captive protective covers.

In the MEASURING (with test specimen) and CALIBRATION operating modes, the

measured values of both the reference pressure sensor and the test item, as well as

their deviation, are displayed in both current pressure units and in %.

In this way the operator is immediately informed whether the test item meets the

accuracy-class or not. The difference between these two modes is that the calibration

data in CALIBRATION mode are stored internally and can later be transferred onto

printable certificates through software (PrintCal or EasyCal).

For the transfer of data to a PC, the CPH6000 ProcessCalibrator has both an RS-232

and a USB connection, selectable via menu.

SETUP menu

Using the SETUP key, one can access the SETUP menu, where the required operating

mode (MEASURING / CALIBRATION / SWITCH-TEST) can be selected and

configured, a stored function can be recalled or a general instrument setting (such as

the menu language) can be changed.

6. Installation and mounting / 7. Commissioning, operation