WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
6.3 Important device settings for calibration using calibration mode
Calibration date
The instrument has an integral real-time clock with date. The current date of a calibration
is stated later in the calibration certificate. Before starting a calibration you must ensure
that the internal date of the CPH6000 is correct (see chapter 7.4.5 "SETUP additional
menu options: CPH-Configuration").
6.4 Unit and resolution
After selecting one of the main menu items (e.g. MEASURING, CALIBRATION or
SWITCH-TEST) from the "SETUP" menu (press "SETUP" key), using the menu
option "Unit", and its associated submenu respectively (move the cursor to "Unit" and
press the right or left arrow), you can set the unit and adjust its resolution (see chapter
7.3 "Operating modes").
Table of available units, including their conversion factors in relation the unit 'bar':
bar
1.00000E+00
mbar
1.00000E-03
hPa
1.00000E-03
psi
6.89475E-02
inHg (0 °C)
3.37690E-02
cmHG (0 °C)
1.33322E-02
MPa
1.00000E+01
kPa
1.00000E-02
Pa
1.00000E-05
mH
2
O (4 °C)
9.80670E-02
cmH
2
O (4 °C)
9.80670E-04
mmH
2
O (4 °C)
9.80670E-05
kg/cm
2
9.80665E-01
inH
2
O (60 °C)
2.48800E-03
mmH
2
O (0 °C)
1.33322E-03
6. Installation and mounting