Installation and mounting – WIKA CPH6000 Benutzerhandbuch

Seite 30

Advertising
background image

30

WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

6.3 Important device settings for calibration using calibration mode
Calibration date

The instrument has an integral real-time clock with date. The current date of a calibration

is stated later in the calibration certificate. Before starting a calibration you must ensure

that the internal date of the CPH6000 is correct (see chapter 7.4.5 "SETUP additional

menu options: CPH-Configuration").

6.4 Unit and resolution

After selecting one of the main menu items (e.g. MEASURING, CALIBRATION or

SWITCH-TEST) from the "SETUP" menu (press "SETUP" key), using the menu

option "Unit", and its associated submenu respectively (move the cursor to "Unit" and

press the right or left arrow), you can set the unit and adjust its resolution (see chapter

7.3 "Operating modes").

Table of available units, including their conversion factors in relation the unit 'bar':

bar

1.00000E+00

mbar

1.00000E-03

hPa

1.00000E-03

psi

6.89475E-02

inHg (0 °C)

3.37690E-02

cmHG (0 °C)

1.33322E-02

MPa

1.00000E+01

kPa

1.00000E-02

Pa

1.00000E-05

mH

2

O (4 °C)

9.80670E-02

cmH

2

O (4 °C)

9.80670E-04

mmH

2

O (4 °C)

9.80670E-05

kg/cm

2

9.80665E-01

inH

2

O (60 °C)

2.48800E-03

mmH

2

O (0 °C)

1.33322E-03