7.4.3 SETUP additional menu items: ReferenceSensor

7. Commissioning, operation

a) Sensor no:

Sensor number of the currently connected CPT6000 reference pressure sensor

b) R.-Start/Range-End:

Start of measuring range and end of measuring range of the currently connected

CPT6000 reference pressure sensor

d) Unit:

Basic pressure unit of the CPT6000 reference pressure sensor

d) Class:

Accuracy of the measuring chain of the CPH6000 with connected CPT6000 refer-

ence pressure sensor

e) Pressuretype:

Pressure type of the currently connected CPT6000 reference pressure sensor

(Gauge/relative pressure or absolute pressure)

f) Overload:

Length of time for which the CPT6000 reference pressure sensor was unaccept-

ably overloaded.

If the value here is not equal to zero, then it is highly probable that

the instrument no longer meets its specified accuracy class. The only

solution for this is an immediate recalibration. (For absolute pressure

sensors < 1 bar this function is deactivated, since for this measuring

range atmospheric pressure already represents an overload)