Gb 4. design and function – WIKA CPH6000 Benutzerhandbuch
Seite 23

23
WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
4. Design and function
4.5 Voltage supply
The internal Lithium-Ion battery, which can be easily charged with the battery charger
supplied with the equipment, serves as the power supply for the instrument.
To charge the CPH6000 rechargeable batteries, the mains plug of the charger/mains
connector must always be plugged in to a mains socket and accessible, so that one can
always remove it from the mains socket without difficulty.
CAUTION!
For EMC reasons, never use the CPH6000 with a mains supply
connected.
The instrument is delivered with a charge level of 25 ... 50 % and should be fully charged
once before being used.
The battery level status (charge in %) is briefly indicated as the equipment is turned
on, and/or it can be viewed during operation via the "SETUP additional menu
options: CPH-Configuration" (see chapter 7.4.5 "SETUP additional menu options:
CPH-Configuration").
Manual quick-release clamp for
easy changeover of the sensors.
Hexagonal anti-rotation stop