WIKA operating instructions ProcessCalibrator, model CPH6000

11069023.04 11/2012 GB/D

GB

SELECT

..Interface

USB

RS232
(8N1)

9600

ON

7.4.6 SETUP additional menu items: Interface

7. Commissioning, operation

This menu allows the configuration of the interface. It can be configured for USB or

RS-232.

For the RS-232 interface there is a choice of 3 different baud rates with fixed 8 used

data bits, no parity and 1 stop bit.

The USB connection can be switched on and off. In order to extend the battery life, the

USB connection should be deactivated when not in use.

7.4.7 SETUP additional menu items: CLEAR CalData
If the SETUP menu item "CLEAR CalData" is selected with the cursor and then the

SELECT button pressed 2x, then all stored calibration data will be deleted and reset.

The status of the deletion process is shown at the left of the lower Info

screen.

Select serial interface

Confirm selection

Change between interface

tape and baud rate