Gb 3. specifications – WIKA DI25 Benutzerhandbuch

Seite 12

Advertising
background image

12

WIKA operating instructions digital indicator, model DI25

14076218.01 12/2013 GB/D

GB

3. Specifications

Serial interface [C5]

The following operations can be executed from an external computer:

Reading and setting up of the different setting values

Reading the measured value and the switching status

Changes of functions

Interface:

based on EIA RS-485

Communication method:

Half-duplex communication, start/stop , synchronous

Transmission rate:

2,400, 4,800, 9,600, 19,200 bps

Parity:

even, odd, no parity

Stop bit:

1, 2

Transmission protocol:

WIKA protocol, MODBUS

RTU, MODBUS

ASCII

Number of connectable

instruments:

max. 31 digital indicators per computer

Detection of transmission

errors:

double error detection by parity and check sum

Transmission protocol

WIKA protocol

MODBUS

ASCII

MODBUS

RTU

Start bit

1

1

1

Data bit

7

7

8

Parity

even

Adjustable (even)

Adjustable (none)

Stop bit

1

Adjustable (1)

Adjustable (1)

The data bit is automatically set up via the choice of the transmission protocol.

Transmitter power supply [P24]
Output voltage:

DC 24 V ± 3 V (with load current 30 mA)

Ripple voltage:

< 200 mV (with load current 30 mA)

Maximum load current: 30 mA

Terminal cover [K]

Screw-on cover of the connection terminals for preventing electric shocks.