Gb 6. commissioning, operation, Connection terminals – WIKA DI25 Benutzerhandbuch
Seite 18
18
WIKA operating instructions digital indicator, model DI25
14076218.01 12/2013 GB/D
GB
6. Commissioning, operation
Connection terminals
Designation
Description
A1, A2, A3
Alarm outputs 1, 2, 3
TRANSMIT OUTPUT
Analogue actual-value output signal
P24
Supply of the transmitter
HOLD
HOLD function
RS-485
Serial interface RS-485
TC
Thermocouple
RTD
Resistance thermometer
DC V
Voltage input
DC mA
Current input (shunt resistor: Measurement shunt 50 Ω included)
6.2.2 Setting up a voltage supply
For digital indicators with AC/DC 24 V voltage supply, it is mandatory to observe the
polarity when using DC currents.
Design
Permissible power supply
Power consumption
AC 100 ... 240 V
AC 85 ... 264 V, 50/60 Hz
approx. 10 VA
AC/DC 24 V
AC/DC 20 ... 28 V, 50/60 Hz
approx. 10 VA
3
2
3
2
1
6
5
4
8
7
A1
NO
3 A
AC 250 V
10
9
A2
NO
3 A
AC 250 V
P24
DC 24 V
30 mA
11
16
15
14
18
17
20
19
8
7
A3
NO
3 A
AC 250 V
RS-485
HOLD
YB (+)
SG
YA (-)
(shunt resistor)
A
A
B
B
DC mA
DC V
RTD
TC
POWER SUPPLY
AC/DC 24 V
POWER SUPPLY
AC 100 ... 240 V
GND
TANSMIT
OUTPUT
Dashed lines show the options