Gb 6. commissioning, operation, Connection terminals – WIKA DI25 Benutzerhandbuch

Seite 18

Advertising
background image

18

WIKA operating instructions digital indicator, model DI25

14076218.01 12/2013 GB/D

GB

6. Commissioning, operation

Connection terminals

Designation

Description

A1, A2, A3

Alarm outputs 1, 2, 3

TRANSMIT OUTPUT

Analogue actual-value output signal

P24

Supply of the transmitter

HOLD

HOLD function

RS-485

Serial interface RS-485

TC

Thermocouple

RTD

Resistance thermometer

DC V

Voltage input

DC mA

Current input (shunt resistor: Measurement shunt 50 Ω included)

6.2.2 Setting up a voltage supply

For digital indicators with AC/DC 24 V voltage supply, it is mandatory to observe the

polarity when using DC currents.

Design

Permissible power supply

Power consumption

AC 100 ... 240 V

AC 85 ... 264 V, 50/60 Hz

approx. 10 VA

AC/DC 24 V

AC/DC 20 ... 28 V, 50/60 Hz

approx. 10 VA

3

2

3

2

1

6

5

4

8

7

A1

NO

3 A

AC 250 V

10

9

A2

NO

3 A

AC 250 V

P24

DC 24 V

30 mA

11

16

15

14

18

17

20

19

8

7

A3

NO

3 A

AC 250 V

RS-485

HOLD

YB (+)

SG

YA (-)

(shunt resistor)

A

A

B

B

DC mA

DC V

RTD

TC

POWER SUPPLY

AC/DC 24 V

POWER SUPPLY

AC 100 ... 240 V

GND

TANSMIT

OUTPUT

Dashed lines show the options

Advertising