Gb 6. commissioning, operation – WIKA DI25 Benutzerhandbuch
Seite 19

19
WIKA operating instructions digital indicator, model DI25
14076218.01 12/2013 GB/D
GB
6. Commissioning, operation
Output signal
Load
4 ... 20 mA
≤ 500 Ω
0 ... 20 mA
≤ 500 Ω
DC 0 ... 1 V
≥ 100 kΩ
DC 0 ... 5 V
≥ 500
DC 1 ... 5 V
≥ 500 kΩ
DC 0 ... 10 V
≥ 1 MΩ
6.3 Functional and operational description
At first, only turn on the voltage supply. The input configuration is indicated on the
display for approximately 3 seconds, the assigned end value is visible on the SV display.
During this time, all outputs and control displays are switched off.
Afterwards, the actual value is shown on the PV display and the alarm value is shown
on the SV display. Now all settings can be carried out in accordance with chapter
6.4 “Parameterisation”.
After parameterisation, switch off voltage supply and wire the sensor, thermocouple or
the resistance thermometer. Voltage supply may only be switched back on after wiring.
With the input configuration 4 ... 20 mA or 0 ... 20 mA, the included 50 Ω measurement
shunt must under all circumstance be connected to the terminals 18 and 19.
Sensor input
°C
°F
PV display
SV display
PV display
SV display
K
J
R
S
B
E
T
N