WIKA operating instructions digital indicator, model DI25
14076218.01 12/2013 GB/D
GB
6. Commissioning, operation
Display
Name, function and setting range
Factory setting
Transmission rate
■
Selection of the transmission rate for communication.
2,400 bps: 24
4,800 bps: 8
9,600 bps: 96
19,200 bps: 92
■
Available only if the serial interface [C5] option is available.
9,600 bps
Parity
■
Setting of parity.
no parity:
even parity:
odd parity:
■
Not available if the serial interface [C5] option is available or if the
WIKA protocol is selected as the transmission protocol.
Stop bit
■
Setting the stop bit
■
Setting range: 1, 2
■
Not available if the serial interface [C5] option is available or if the
WIKA protocol is selected as the transmission protocol.
1
6.4.2 Setting the alarm value
To configure the alarm values, press MODE key. To switch to the next parameter, press
the MODE key once again. After setting up the final parameter, exit the menu by press-
ing the MODE key.
Display
Name, function and setting range
Factory setting
Alarm value alarm 1 (A1)
■
Entering the alarm value for alarm output A1.
■
Not available if no alarm type is selected for the alarm A1.
■
See table “setting ranges alarm types”
0 °C
Alarm value alarm 2 (A2)
■
Entering the alarm value for alarm output A2.
■
Not available when the option transmitter power supply [P24] is
available or if no alarm type has been selected for alarm A2.
■
See table “setting ranges alarm types”
0 °C
Alarm value alarm 3 (A3)
■
Entering the alarm behaviour for alarm output A3.
■
Not available if no alarm type is selected for the alarm A3 or if the
high/low alarm was selected.
■
See table “setting ranges alarm types”
0 °C