Gb 6. commissioning, operation – WIKA CTM9100-150 Benutzerhandbuch

Seite 21

Advertising
background image

21

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

6. Commissioning, operation

6.6 Testing of temperature sensors

To test temperature sensors, connect a separate tempera-

ture-measuring instrument to the test item. By comparing the

temperature displayed on the external measuring instrument

with the reference temperature, there is evidence of the

status of the test item. This ensured that the test item requires

a short time until it reaches the temperature of the metal

block or liquid bath.

WARNING!

There is no possibility for calibrating earthed

thermocouples, since the heating block is

earthed and this would therefore lead to

false measurement results.

6.7 Start-up procedure

If the calibrator is not used for a long time, because of the

materials used (magnesium oxide), moisture can penetrate

the heating elements.

Following transport or storage of the calibrator in moist

environments, the heating elements must be taken to higher

temperatures slowly.

During the drying process,it should be assumed that the

calibrator has not yet reached the required insulation voltage

for protection class I. The start-up set point is T

anf

= 120 °C

with a holding time of T

n

= 15 min.

6.8 Switching on the calibrator/micro calibration bath

1. Connect to the mains using the mains connector supplied.

2. Switch on the mains switch.
The controller will be initialised.

The upper PV display will show tEst.

In the lower SV display, the version number (e.g. rL 2.2) will

be shown.

After approx. 5 secs, the initialisation will be complete and

the calibration mode will automatically be displayed.

The built-in heating or cooling elements will temper the metal

block automatically from room temperature to the controller’s

set temperature.

6.9 Reference and set temperature display
Upper PV display:

The red 4-digit, 7-segment display shows the current temper-

ature of the metal block or liquid bath.

Lower SV display:

The green, 4-digit, 7-segment display shows the current set

temperature of the metal block or the liquid bath.

Once the set temperature has been reached, through short

switch impulses, the radiated heating energy from the metal

block or liquid bath is delivered, so that the temperature

within remains constant.

Reference and set temperature display

6.10 Control of the reference temperature

The red LED OUT 1 indicates that the heater is turned on.

During the heating phase a steady light indicates that heating

energy is being supplied. A flashing LED indicates that the

reference temperature (set temperature) has almost been

reached and therefore the heating energy is only being

supplied in short bursts.

Display with LED OUT 1

To ensure good temperature stability, the cycle time of the

controller is set low and the control output is usually raised.

Control achieved via PID algorithm

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: