Gb 9. maintenance, cleaning and recalibration, Maintenance, cleaning and recalibration – WIKA CTM9100-150 Benutzerhandbuch

Seite 34

Advertising
background image

34

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

9. Maintenance, cleaning and recalibration

9. Maintenance, cleaning and recalibration

9.1 Maintenance

The instruments described here are maintenance-free.

Repairs must only be carried out by the manufacturer.

This does not apply to the fuse replacement. Before replac-

ing the fuse, the calibrator or the micro calibration bath must

be disconnected by unplugging the mains cable from the

power supply outlet.

9.2 Cleaning

CAUTION!

Cool down the calibrator or micro calibra-

tion bath as described in chapter 8

“Cooling the metal block or micro calibra-

tion bath”.

Prior to cleaning, switch off and discon-

nect the calibrator or micro calibration

bath from the mains.

Clean the instrument with a moist cloth.

Electrical connections must not come into

contact with moisture.

Wash or clean the dismounted instrument

before returning it, in order to protect

persons and the environment from

exposure to residual media.

Residual media in the dismounted instru-

ment can result in a risk to persons, the

environment and equipment. Take suffi-

cient precautionary measures.

For information on returning the instrument

see chapter 11.2 “Return”.

9.2.1 Cleaning calibrators with inserts

With calibrators with inserts, during operation, a small

amount of metallic dust can cause the block and the sleeve

to become jammed. To prevent this, on a regular basis and

before any long period out-of-use, remove the inserts from

the calibrator heating block. Blow out heating block bores

with compressed air and clean the bore and sleeve with a dry

cloth.

9.2.2 Cleaning fan guards

On the base of each calibrator is a dense fan grille, through

which the cooling air is supplied to the calibrator. Depending

on the cleanliness of the air, clean the grille at regular inter-

vals by vacuuming or brushing.

9.2.3 Cleaning the micro calibration bath

Remove as much of the silicone oil as possible from the tank.

Then take the sensor basket out of the tank and clean the

basket, magnetic stirrer and tank with water, in which plenty

of detergent has been added. Let everything dry thoroughly.

When distilled water has been used, remove the calibration

liquid and let the sensor basket, magnetic stirrer and tank dry

thoroughly.

9.2.4 External cleaning

Clean the outside of the calibrator or micro calibration bath

with a damp cloth and some water, or with a solvent-free light

detergent.

9.3 Recalibration
DKD/DAkkS certificate

The calibrator or the micro calibration bath has been adjust-

ed and tested before delivery using measuring instruments

that are traceable to nationally-recognised standards.

On the basis of DIN ISO 10012, the calibrator or the micro

calibration bath, depending on the application, should be

verified at appropriate periodic intervals.

We recommend that the instrument is regularly recalibrated

by the manufacturer, with time intervals of approx. 12 months

or approximately 500 hours of operation. Every factory

recalibration includes, additionally, an extensive free-of-

charge check of all system parameters with respect to their

compliance with the specification. The basic settings will be

corrected if necessary.

The basis of the recalibration is the guidelines of the German

Calibration Service, DKD R5-4. The measures described

here should be used and followed for recalibration.

Advertising
Dieses Handbuch ist für die folgenden Produkte bezogen werden: