WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

13. Heating and cooling times

13.5 Dry-well temperature calibrator, model CTD9100-650
Measuring conditions

■

All times refer to a Pt100 reference sensor with Ø 6 mm

■

The reference sensor is positioned, at full depth, centrally

in the insert

■

All times are transition times and do not account for the

required settling time

■

The measurements were carried out at a room tempera-

ture of approx. 23 °C

Insert

Heating up:

Time

Cooling:

Time

25 °C to 40 °C

0:54 min

650 °C to 600 °C

2:25 min

40 °C to 50 °C

0:22 min

600 °C to 550 °C

2:33 min

50 °C to 100 °C

1:18 min

550 °C to 500 °C

2:55 min

100 °C to 150 °C

1:06 min

500 °C to 450 °C

3:27 min

150 °C to 200 °C

1:03 min

450 °C to 400 °C

4:01 min

200 °C to 250 °C

1:05 min

400 °C to 350 °C

4:39 min

250 °C to 300 °C

1:06 min

350 °C to 300 °C

5:36 min

300 °C to 350 °C

1:09 min

300 °C to 250 °C

6:46 min

350 °C to 400 °C

1:21 min

250 °C to 200 °C

8:32 min

400 °C to 450 °C

1:30 min

200 °C to 150 °C

11:22 min

450 °C to 500 °C

1:32 min

150 °C to 100 °C

17:01 min

500 °C to 550 °C

1:38 min

100 °C to 50 °C

52:37 min

550 °C to 600 °C

1:55 min

50 °C to 40 °C

15:23 min

600 °C to 650 °C

2:33 min

40 °C to 25 °C

1:01:58 min