13. Heating and cooling times

13.5 Dry-well temperature calibrator, model CTD9100-650
Measuring conditions

All times refer to a Pt100 reference sensor with Ø 6 mm

The reference sensor is positioned, at full depth, centrally

in the insert

All times are transition times and do not account for the

required settling time

The measurements were carried out at a room tempera-

ture of approx. 23 °C

Insert

Heating up:

Time

Cooling:

Time

25 °C to 40 °C

0:54 min

650 °C to 600 °C

2:25 min

40 °C to 50 °C

0:22 min

600 °C to 550 °C

2:33 min

50 °C to 100 °C

1:18 min

550 °C to 500 °C

2:55 min

100 °C to 150 °C

1:06 min

500 °C to 450 °C

3:27 min

150 °C to 200 °C

1:03 min

450 °C to 400 °C

4:01 min

200 °C to 250 °C

1:05 min

400 °C to 350 °C

4:39 min

250 °C to 300 °C

1:06 min

350 °C to 300 °C

5:36 min

300 °C to 350 °C

1:09 min

300 °C to 250 °C

6:46 min

350 °C to 400 °C

1:21 min

250 °C to 200 °C

8:32 min

400 °C to 450 °C

1:30 min

200 °C to 150 °C

11:22 min

450 °C to 500 °C

1:32 min

150 °C to 100 °C

17:01 min

500 °C to 550 °C

1:38 min

100 °C to 50 °C

52:37 min

550 °C to 600 °C

1:55 min

50 °C to 40 °C

15:23 min

600 °C to 650 °C

2:33 min

40 °C to 25 °C

1:01:58 min