13. Heating and cooling times
13.5 Dry-well temperature calibrator, model CTD9100-650
Measuring conditions
■
All times refer to a Pt100 reference sensor with Ø 6 mm
■
The reference sensor is positioned, at full depth, centrally
in the insert
■
All times are transition times and do not account for the
required settling time
■
The measurements were carried out at a room tempera-
ture of approx. 23 °C
Insert
Heating up:
Time
Cooling:
Time
25 °C to 40 °C
0:54 min
650 °C to 600 °C
2:25 min
40 °C to 50 °C
0:22 min
600 °C to 550 °C
2:33 min
50 °C to 100 °C
1:18 min
550 °C to 500 °C
2:55 min
100 °C to 150 °C
1:06 min
500 °C to 450 °C
3:27 min
150 °C to 200 °C
1:03 min
450 °C to 400 °C
4:01 min
200 °C to 250 °C
1:05 min
400 °C to 350 °C
4:39 min
250 °C to 300 °C
1:06 min
350 °C to 300 °C
5:36 min
300 °C to 350 °C
1:09 min
300 °C to 250 °C
6:46 min
350 °C to 400 °C
1:21 min
250 °C to 200 °C
8:32 min
400 °C to 450 °C
1:30 min
200 °C to 150 °C
11:22 min
450 °C to 500 °C
1:32 min
150 °C to 100 °C
17:01 min
500 °C to 550 °C
1:38 min
100 °C to 50 °C
52:37 min
550 °C to 600 °C
1:55 min
50 °C to 40 °C
15:23 min
600 °C to 650 °C
2:33 min
40 °C to 25 °C
1:01:58 min