13. Heating and cooling times

Infra-red insert

Heating up:

Time

Cooling:

Time

-20 °C to -15 °C

0:30 min

150 °C to 125 °C

2:26 min

-15 °C to -10 °C

0:30 min

125 °C to 100 °C

2:52 min

-10 °C to 0 °C

1:00 min

100 °C to 75 °C

3:36 min

0 °C to 25 °C

2:42 min

75 °C to 50 °C

4:27 min

25 °C to 50 °C

3:06 min

50 °C to 25 °C

6:03 min

50 °C to 75 °C

3:26 min

25 °C to 0 °C

8:59 min

75 °C to 100 °C

4:12 min

0 °C to -10 °C

5:33 min

100 °C to 125 °C

5:38 min

-10 °C to -15 °C

3:31 min

125 °C to 150 °C

9:49 min

-15 °C to -20 °C

4:32 min

13.11 Model CTM9100-150 as surface temperature calibrator
Measuring conditions

■

All times refer to a Pt100 reference sensor with Ø 3 mm,

L = 300 mm

■

The reference sensor is inserted to a depth of 111 mm

■

All times are transition times and do not account for the

required settling time

■

The measurements were carried out at a room tempera-

ture of approx. 23 °C

■

It is controlled by an external reference sensor (3 x 300 mm)

■

Without cap, without covering

13.10 Model CTM9100-150 as an infrared black body source
Measuring conditions

Surface insert

Heating up:

Time

Cooling:

Time

-20 °C to -15 °C

0:46 min

150 °C to 125 °C

3:11 min

-15 °C to -10 °C

0:45 min

125 °C to 100 °C

3:17 min

-10 °C to 0 °C

1:15 min

100 °C to 75 °C

3:51 min

0 °C to 25 °C

2:57 min

75 °C to 50 °C

5:02 min

25 °C to 50 °C

3:16 min

50 °C to 25 °C

6:58 min

50 °C to 75 °C

3:37 min

25 °C to 0 °C

11:55 min

75 °C to 100 °C

4:46 min

0 °C to -10 °C

9:19 min

100 °C to 125 °C

6:18 min

-10 °C to -15 °C

7:44 min

125 °C to 150 °C

10:45 min

-15 °C to -20 °C

12:35 min

■

All times refer to a Pt100 reference sensor with Ø 3 mm,

L = 150 mm

■

The reference sensor is inserted to a depth of 51 mm, and

is located under the front surface

■

All times are transition times and do not account for the

required settling time

■

The measurements were carried out at a room tempera-

ture of approx. 23 °C

■

It is controlled by an external reference sensor (3 x 300 mm)

■

Without cap, without covering