WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements ... / 8. Cooling down the metal blocks or ...

Input of the dwell time

Press button P.

On the upper PV display, dur.t will blink.

On the lower SV display, the corresponding currently-select-

ed dwell time is displayed.

By pressing button P, the newly-set dwell time, dur.t, is

accepted.

The screen returns to the parameter level.

CAUTION!

If no button has been pressed for approx.

15 secs., an automatic return to one level

above in Calibration mode occurs.

After the setting has been made, the calibra-

tor or the micro calibration bath will not use

the new values until you either change the

set temperature or the calibrator or the micro

calibration bath has been turned off and

then on again.

8. Cooling down the metal blocks or micro calibration baths

WARNING! Risk of burns!

Before transporting or touching the metal

block or micro calibration bath, make sure

that it has cooled down sufficiently, as there

is an acute risk of burns from both the metal

block and micro calibration bath and also

from the test item.

In order that the calibrator or the micro

calibration bath can be brought quickly from

a higher to a lower temperature, set the set

temperature to a lower temperature (e.g.

room temperature).

With a heating instrument, the built-in ventilator will automati-

cally slowly switch to higher speed, which will provide a

cooling airflow.

The LED OUT 2 signals the status of the outputs for the

fan control. If LED OUT 2 is on, the fan is running at higher

speed, if LED OUT 2 is not on, the fan is running at lower

speed.

With a heating and cooling instrument, the controller will

switch the active cooling on. The LED OUT 2 signals the

status of the outputs for the fan control. If LED OUT 2 is on,

the active cooling is running, if LED OUT 2 is not on, the

cooling is not active.

WARNING!

After switching off or removing the mains

connection, no cooling air will be provided

by the built-in fan. A sufficient thermal

decoupling between the metal block and

liquid bath and body is nevertheless guaran-

teed.

By pressing button 5, the dwell time, dur.t,is increased.

By pressing button 6, the dwell time, dur.t,is decreased.

By pressing button 5 or 6 the value is

increased or decreased by 0.1. If the keys

are held down for at least 1 sec., however,

the value increases or decreases more

rapidly, and after 2 secs. even faster, so

that the desired value can be reached very

quickly.