WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
13. Heating and cooling times
13.6 Micro calibration bath, model CTB9100-165
Measuring conditions
■
All times refer to a Pt100 reference sensor with Ø 6 mm
■
The reference sensor is positioned centrally in the tank,
5 mm over the strainer insert
■
All times are transition times and do not account for the
required settling time
■
The measurements were carried out at a room tempera-
ture of approx. 23 °C, with the cover off the bath
Distilled water
Heating up:
Time
Cooling:
Time
2 °C to 25 °C
5:31 min
90 °C to 75 °C
3:09 min
25 °C to 50 °C
6:49 min
75 °C to 50 °C
7:06 min
50 °C to 75 °C
8:07 min
50 °C to 25 °C
10:18 min
75 °C to 90 °C
6:19 min
25 °C to 2 °C
14:52 min
Silicone oil 5 CS
Heating up:
Time
Cooling:
Time
-30 °C to -25 °C
0:56 min
120 °C to 100 °C
32:24 min
-25 °C to -15 °C
1:06 min
100 °C to 75 °C
3:40 min
-15 °C to 0 °C
1:18 min
75 °C to 50 °C
4:48 min
0 °C to 25 °C
2:46 min
50 °C to 25 °C
6:41 min
25 °C to 50 °C
2:37 min
25 °C to 0 °C
8:50 min
50 °C to 75 °C
3:10 min
0 °C to -15 °C
10:36 min
75 °C to 100 °C
4:23 min
-15 °C to -25 °C
15:01 min
100 °C to 120 °C
5:05 min
-25 °C to -30 °C
23:19 min
Silicone oil 10 CS
Heating up:
Time
Cooling:
Time
-30 °C to -25 °C
1:17 min
165 °C to 150 °C
1:54 min
-25 °C to -15 °C
1:17 min
150 °C to 125 °C
2:37 min
-15 °C to 0 °C
1:20 min
125 °C to 100 °C
3:11 min
0 °C to 25 °C
1:56 min
100 °C to 75 °C
3:59 min
25 °C to 50 °C
2:30 min
75 °C to 50 °C
5:02 min
50 °C to 75 °C
3:13 min
50 °C to 25 °C
6:57 min
75 °C to 100 °C
4:24 min
25 °C to 0 °C
8:26 min
100 °C to 125 °C
6:47 min
0 °C to -15 °C
9:58 min
125 °C to 150 °C
12:51 min
-15 °C to -25 °C
15:33 min
150 °C to 165 °C
18:21 min
-25 °C to -30 °C
29:45 min