Gb 3. specifications – WIKA CTM9100-150 Benutzerhandbuch
Seite 9

9
WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
3. Specifications
Approvals and certificates, CTD9100 series
CE conformity
EMC directive
2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity
(industrial application)
Low voltage directive
2006/95/EC, EN 61010-1, safety requirements for electrical equipment for measure-
ment, control and laboratory use
Certificate
Calibration
3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
For further specifications see WIKA data sheet CT 41.28 and the order documentation.
3.2 Micro calibration bath, CTB9100 series
Model CTB9100-165
Model CTB9100-225
Temperature range
-35 ... +165 °C
t
amb.
... 225 °C, optional t
amb.
... 255 °C
Accuracy
±0.2 K
±0.3 K
Stability
±0.05 K
±0.05 K
Display resolution
0.1 °C
0.1 °C
Heating time incl. stabilisation
2)
30 min from 20 to 160 °C
20 min from 20 to 225 °C
Cooling time incl. stabilisation
2)
30 min from +20 to -20 °C
30 min from 225 to 50 °C
Immersion depth
150 mm
150 mm
Volume
Approx. 0.6 litres
Approx. 0.6 litres
Tank dimensions
Ø 60 x 165 mm
Ø 60 x 165 mm
Voltage supply
Power supply
AC 100 ... 240 V, 50/60 Hz
AC 230/240 V, 50/60 Hz
(AC 100/115 V, 50/60 Hz)
4)
Power consumption
375 VA
1,000 VA
Power cord
for Europe, AC 230 V
for Europe, AC 230 V
Communication
Interface
RS-485
RS-485
Case
Dimensions
215 x 305 x 425 mm (W x H x D)
150 x 270 x 400 mm (W x H x D)
Weight
12 kg
7.9 kg
2) The reference thermometer with which the measurements are performed, has a diameter of 6 mm.
4) AC 115 V power supply must be specified on the order, otherwise an AC 230 V one will be delivered.
Accessories
Model CTB9100-165
Model CTB9100-225
DC 200.05 silicone oil: -40 ... +130 °C,
FP* = 133 °C
from -35 ... +130 °C, very well usable
not recommended
DC 200.10 silicone oil: -35 ... +160 °C,
FP* = 163 °C
from -35 ... +160 °C, well usable
not recommended
DC 200.20 silicone oil: 10 ... 220 °C,
FP* = 232 °C
not recommended
from 40 ... 225 °C, well usable
DC 200.50 silicone oil: 25 ... 250 °C,
FP* = 280 °C
not recommended
from 80 ... 255 °C, well usable
* FP = Flash point