WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
7. Operating elements of the calibrator/micro calibration bath
7.4.4 Switching on manual control
Switch off the automatic control of the calibrator or micro
calibration bath, and reach the required temperature with
manual control.
Press button P for approx. 5 secs., this opens the main menu.
The upper PV display will show OPEr.
In the lower SV display, the LED SET will blink.
Press button 5 or button 6 until OPLO is displayed.
The upper PV display will show OPLO.
In the lower SV display, the LED SET will blink.
Confirm by pressing the button P.
On the upper PV display, the current reference temperature
is indicated.
On the lower SV display, an H and the current set output
power in % is displayed.
By pressing button 5, the output power is increased.
By pressing button 6, the output power is decreased.
CAUTION!
By pressing button 5 or 6 the value is
increased or decreased by 0.1 %. If the keys
are held down for at least 1 sec., however,
the value increases or decreases more
rapidly, and after 2 secs. even faster, so
that the desired value can be reached very
quickly.
7.4.5 Switching off manual control
The manual control is switched on if the following display is
shown:
On the upper PV display, the current reference temperature
is indicated.
On the lower SV display, an H and the current set output
power in % is displayed.
The manual control is switched back off by pressing button P
for approx. 5 secs., this opens the main menu.
The upper PV display will show ’rEG.
In the lower SV display, the LED SET will blink.
Confirm switching the automatic control on by pressing
button P.
Display in main menu
Manual control OPLO menu
Display with setting manual control OPLO
Display in main menu
Display with setting manual control OPLO