7. Operating elements of the calibrator/micro calibration bath

7.4.4 Switching on manual control

Switch off the automatic control of the calibrator or micro

calibration bath, and reach the required temperature with

manual control.

Press button P for approx. 5 secs., this opens the main menu.

The upper PV display will show OPEr.

In the lower SV display, the LED SET will blink.

Press button 5 or button 6 until OPLO is displayed.

The upper PV display will show OPLO.

In the lower SV display, the LED SET will blink.

Confirm by pressing the button P.

On the upper PV display, the current reference temperature

is indicated.

On the lower SV display, an H and the current set output

power in % is displayed.

By pressing button 5, the output power is increased.

By pressing button 6, the output power is decreased.

CAUTION!

By pressing button 5 or 6 the value is

increased or decreased by 0.1 %. If the keys

are held down for at least 1 sec., however,

the value increases or decreases more

rapidly, and after 2 secs. even faster, so

that the desired value can be reached very

quickly.

7.4.5 Switching off manual control

The manual control is switched on if the following display is

shown:

On the upper PV display, the current reference temperature

is indicated.

On the lower SV display, an H and the current set output

power in % is displayed.

The manual control is switched back off by pressing button P

for approx. 5 secs., this opens the main menu.

The upper PV display will show ’rEG.

In the lower SV display, the LED SET will blink.

Confirm switching the automatic control on by pressing

button P.

Display in main menu

Manual control OPLO menu

Display with setting manual control OPLO

Display in main menu

Display with setting manual control OPLO