WIKA CTM9100-150 Benutzerhandbuch
Seite 32

32
WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
7. Operating elements of the calibrator/micro calibration bath
Select the parameter SLoF with button 5 or button 6.
The upper PV display will show 'rEG.
In the lower SV display, SLoF will blink.
Press button P.
On the upper PV display, SLoF will blink.
On the lower SV display, the corresponding currently-
selected cooling gradient is displayed.
By pressing button 5, the cooling gradient, SLoF, will be
increased.
By pressing button 6, the cooling gradient, SLoF, will be
decreased.
By pressing button 5 or 6 the value is
increased or decreased by 0.1. If the keys
are held down for at least 1 sec., however,
the value increases or decreases more
rapidly, and after 2 secs. even faster, so
that the desired value can be reached very
quickly.
By pressing button P, the newly-set cooling
gradient, SLoF,is accepted.
The display returns to the parameter level
and the other parameters can be set.
CAUTION!
If no button has been pressed for approx.
15 secs., an automatic return to one level
above in calibration mode occurs.
After the setting has been made, the calibra-
tor or the micro calibration bath will not use
the new values until you either change the
set temperature or the calibrator or the micro
calibration bath has been turned off and
then on again.
7.4.8.3 Setting the dwell time
The dwell time, dur.t, is active when the set temperature,
SP1, has been reached. Subsequently, the calibrator or
micro calibration bath automatically switches to set tempera-
ture SP2.
The setting range stretches from 99:59 [hh:min] to
0:00 [hh:min].
CAUTION!
The function is deactivated if dur.t = InF (in
no Function) is set.
You are at the parameter level (as described in chapter
7.4.1 “Menu structure, parameter levels”).
The upper PV display will show 'rEG.
In the lower SV display, SLor will blink.
Select the parameter dur.t with button 5 or button 6.
The upper PV display will show 'rEG.
In the lower SV display, dur.t will blink.
Input of the cooling gradient
Display on cooling gradient input
Parameter for heating gradient, SLor
Parameter for dwell time, dur.t