WIKA CTM9100-150 Benutzerhandbuch

Seite 31

Advertising
background image

31

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements of the calibrator/micro calibration bath

If button P is pressed again, this opens the parameter

level.

The upper PV display will show 'rEG.

In the lower SV display, SLor will blink.

7.4.8.1 Setting the heating gradient

The heating gradient, SLor, is active if the reference temper-

ature is lower than the set temperature.

The setting range stretches from 99.99 °C/min to 0.00 °C/min.

CAUTION!

The function is deactivated if SLor = InF (in

no Function) is set.

You are at the parameter level (as described in chapter

7.4.1 “Menu structure, parameter levels”).

The upper PV display will show 'rEG.

In the lower SV display, SLor will blink.

Press button P.

On the upper PV display, SLor will blink.

On the lower SV display, the corresponding currently-

selected heating gradient is displayed.

By pressing button 5, the heating gradient, SLor, will be

increased.

By pressing button 6, the heating gradient, SLor, will be

decreased.

By pressing button 5 or 6 the value is

increased or decreased by 0.1. If the keys

are held down for at least 1 sec., however,

the value increases or decreases more

rapidly, and after 2 secs. even faster, so

that the desired value can be reached very

quickly.

By pressing button P, the newly-set heating gradient, SLor,

is accepted.

The display returns to the parameter level and the other

parameters can be set.

CAUTION!

If no button has been pressed for approx.

15 secs., an automatic return to one level

above in calibration mode occurs.

After the setting has been made, the calibra-

tor or the micro calibration bath will not use

the new values until you either change the

set temperature or the calibrator or the micro

calibration bath has been turned off and

then on again.

7.4.8.2 Setting the cooling gradient

The cooling gradient, SLoF, is active if the reference temper-

ature is higher than the set temperature.

The setting range stretches from 99.99 °C/min to 0.00 °C/min.

CAUTION!

The function is deactivated if SLoF = InF (in

no Function) is set.

You are at the parameter level (as described in chapter

7.4.1 “Menu structure, parameter levels”).

The upper PV display will show 'rEG.

In the lower SV display, SLor will blink.

Input of the heating gradient

Parameter for heating gradient, SLor

Parameter for heating gradient, SLor

Parameter for heating gradient, SLor