WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements of the calibrator/micro calibration bath

7.4.6 Setting and storing fixed set temperatures

In order to store the set temperature in the calibrator or micro

calibration bath, the appropriate set-point memory must be

opened.

In calibration mode, press button P for approx. 5 secs., this

opens the main menu.

The upper PV display will show OPEr.

In the lower SV display, the LED SET will blink.

If button P is pressed again, this opens the groups level.

The upper PV display will show OPEr.

In the lower SV display, ’SP is shown and additionally LED

SET will blink.

If button P is pressed again, this opens the parameter level.

The upper PV display will show ’SP.

In the lower SV display, the set-point memory SP1 and

additionally LED SET will blink.

Select one of the four set-point memories SP1, SP2, SP3 or

SP4 using button 5 or 6.

By pressing button P the corresponding set-point memory

will then be opened.

In the upper PV display, the selected set-point memory e.g.

SP3 will blink.

On the lower SV display, the corresponding currently-select-

ed set temperature is displayed.

By pressing button 5, the set temperature is increased.

By pressing button 6, the set temperature is decreased.

By pressing button 5 or 6 the value is

increased or decreased by 0.01 °C. If the

keys are held down for at least 1 sec.,

however, the value increases or decreases

more rapidly, and after 2 secs. even faster,

so that the desired value can be reached

very quickly.

By pressing button P, the newly-set temperature is accepted.

The set-point memory is left and the screen returns to the

parameter level.

To return to calibration mode, press button 6 or button 5

for a long time.

If no button has been pressed for approx.

15 secs., an automatic return to one level

above in calibration mode occurs.

’SP Group

Operator menu OPEr

Input in set-point memory SP3

Parameter for set-point memory SP1