29

WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements of the calibrator/micro calibration bath

7.4.7 Recalling the stored set temperatures

The set temperature can be called up from calibration mode.

Press button U for approx. 2 secs. The current set-point

memory will be opened.

On the upper PV display, the current reference temperature

is indicated.

On the lower SV display, the set-point memory (SP1, SP2,

SP3 or SP4) is shown for 2 secs., and then the correspond-

ing currently-selected set temperature is displayed.

It initially displays the set-point memory (SP1, SP2, SP3 or

SP4).

Then it displays the stored set temperature.

In order to retrieve a different stored set point, press button U

again.

The selected temperature value will be immediately adopted

and approached.

7.4.8 Setting the ramp control and a temperature profile

Using the ramp control, the time can be defined over which

the target temperature is achieved. This time may be shorter

or longer than that which the calibrator or the micro calibra-

tion bath would normally require.

By changing the set temperature or turning on the calibrator

or micro calibration bath, it is automatically determined which

of the gradients to use (heating gradient, SLor, or cooling

gradient, SLoF).

In the calibrator or in the micro calibration bath, the dwell

time, dur.t, can be programmed so that the set temperature

reached is automatically switched from the set-point memory

SP1 to the set temperature in set-point memory SP2 after a

predetermined time.

A simple temperature profile can be generated.

After switching on the calibrator or micro calibration bath, the

temperature profile will run automatically.

Heating gradient, SLor

The heating gradient, SLor, is active if the reference temper-

ature is lower than the set temperature.

Each calibrator model has a maximum heating output, and

thus only settings < than this heating power make sense and

actually increase the time taken to achieve a set temperature.

Calibrator model

(heating/cooling)

Setting for

SLor

CTD9100-165

< 7 °C/min

CTB9100-165 / CTM9100-150

with silicone oil 10 CS

< 3 °C/min

CTB9100-165 / CTM9100-150

with distilled water

< 5 °C/min

CTM9100-150 as calibrator

< 3 °C/min

CTM9100-150 as infra-red calibrator < 3 °C/min
CTM9100-150 as surface calibrator

< 3 °C/min

Calibrator model

(heating)

Setting for

SLor

CTD9100-450 / CTD9100-650

< 35 °C/min

CTB9100-225

with silicone oil 20 CS

< 22 °C/min

CTB9100-225

with distilled water

< 12 °C/min

Ramp control and temperature profile

Display on calling the set temperature