WIKA operating instructions, models CTD9100, CTB9100, CTM9100

11263911.02 05/2014 GB/D

GB

7. Operating elements of the calibrator/micro calibration bath

7. Operating elements of the calibrator/micro calibration bath

There are three modes available for operation
Calibration mode

In this standard operating state the calibration of test items

can be made.

Set-point mode

In this mode, the set temperature is entered.

Main menu

In this mode, all settings, such as specifying the desired

temperature or setting the control parameters are made.

7.1 Selection of the operating modes for the CTM9100-150

To operate the multi-function calibrator in the desired mode

of operation, first specify the correct linearisation for the

controller. In the main menu (via the P button) select the

desired function using the arrow keys.

LI Operates as a micro calibration bath

DB Operates as a dry block

Ir

Operates as an infrared black body source

SU Operates as a surface temperature calibrator

The desired function is accepted using the P button.

On the main screen, the desired linearisation (LI, DB, Ir or

SU) is displayed in the PV line every 5 secs.

7.1.1 Method of operation in calibration mode within

individual operating modes

Micro calibration bath operating mode (possible with or

without insert for liquids)

Insert the magnetic stirrer and sensor basket.

Fill the micro calibration bath (see chapter 6.3.2 “Filling

the micro calibration bath”).

Set the speed of the magnetic stirrer, in order to achieve

the best homogeneity possible.

In order to ensure the correct linearisation, set LI in the

controller. For this, press the P button for approx. 5 secs.

and confirm the appropriate setting in the main menu with P.

Set the switch to the left of the controller to internal

reference.

Angled thermometers, large-diameter thermometers or

thermometers with special designs cannot be calibrated

with a dry-well calibrator. For this reason the multi-function

calibrator also has the possibility to function as a stirred liquid

bath. The liquid is circulated using a magnetic stirrer and thus

provides very good temperature distribution within the bath.

Select the liquid used depending upon the desired calibration

temperature.

Dry-block operating mode

Cleaning the tank (if needed)

Inserting the insert (aluminium)

Set the speed of the magnetic stirrer to “0”.

In order to ensure the correct linearisation, set DB in the

controller. For this, press the P button for approx. 5 secs.

and confirm the appropriate setting in the main menu with P.

Set the switch to the left of the controller to external

reference. Connect the delivered external reference in to

the socket provided and slide into the appropriate bore in

the insert.

The insert has several bores into which the thermometer

probes being calibrated and the external reference, for

comparative calibration, can be inserted. The block is either

heated or cooled to the desired calibration temperature.

Once a stable temperature has been reached, the

temperature probes to be calibrated can be compared with

the reference thermometer.

Infra-red operating mode

Cleaning the tank (if needed)

Insert the insert (hollow, ceramic coated)

Set the speed of the magnetic stirrer to “0”.

In order to ensure the correct linearisation, set Ir in the

controller. For this, press the P button for approx. 5 secs.

and confirm the appropriate setting in the main menu with P.

Set the switch to the left of the controller to external

reference. Connect the delivered external reference into

the socket provided and slide it into into the appropriate

bore on the outer edge of the insert.

The measuring spot of the pyrometer being calibrated must

be smaller than the diameter of the infrared insert.

Surface operating mode

Cleaning the tank (if needed)

Insert the insert (hollow, fitted with a collar at the top)

Set the speed of the magnetic stirrer to “0”.

In order to ensure the correct linearisation, set SU in the

controller. For this, press the P button for approx. 5 secs.

and confirm the appropriate setting in the main menu with P.

Set the switch to the left of the controller to external

reference. Connect the delivered external reference in to

the socket provided and slide into the appropriate bore

directly under the surface of the insert.