10. Faults

10. Faults

Error

Causes

Measures

- - - -

Break in the internal reference sensor or the

internal reference sensor is defective.

Send the instrument back to the manufac-

turer or service partner for repair.

uuuu

Measured temperature under the limit value

of the internal reference sensor (Under-

range -200 °C)

oooo

Measured temperature over the limit value

of the internal reference sensor (Overrange

+850 °C)

ErEP

Possible error in the controller's EEPROM

memory.

Press button P

Fan is not running

Fan is defective or blocked.

The temperature switch may have been

triggered and the current supply to the

heating elements switched off.

Send the instrument back to the manufac-

turer or service partner for repair.

Final temperature was not

reached

Solid state relay defective, or the heating-

cooling element has a short-circuit or has

aged.

Send the instrument back to the manufac-

turer or service partner for repair.

No display

The controller is defective.

Send the instrument back to the manufac-

turer or service partner for repair.

No function

The mains connection is not made properly

or the fuse is defective.

Check the mains connection and fuse.

CAUTION!

If faults cannot be eliminated by means of the measures listed above, the instrument must be shut down

immediately.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter 11.2 “Return”.