Completion of the restoration – Ivoclar Vivadent IPS e.max CAD Chairside v.1 Benutzerhandbuch

Seite 89

Advertising
background image

37

Completion of the Restoration

Once the IPS e.max CAD restoration has cooled to room temperature, proceed with the following steps:

– Remove the restoration from the hardened auxiliary firing paste.

– Remove any residue with ultrasound in a water bath and/or with steam.

Do not blast the restoration with Al

2

O

3

or glass polishing beads.

– Final try-in of the restoration before cementation.

– If adjustments by grinding are required, make sure that no overheating of glass-ceramic occurs.

– Use fine-grained diamonds (< 60 µm) for finishing, followed by rubber polishers at low speed and limited pressure. For

the selection of the appropriate grinding tools, refer to the Ivoclar Vivadent Flow Chart “Recommended grinding tools –

chairside”.

– Finally, polish the adjusted areas to a high gloss (e.g. using OptraFine).

– Thoroughly clean the restoration.

Remove the cool restoration from the hardened IPS Object Fix Putty or Flow

Remove residue with ultrasound in a water bath…

... or with steam.

Final try-in of the restoration before cementation

Staining

Technique