ZOLL E Series Monitor Defibrillator Rev H Benutzerhandbuch
Seite 128

E Series Bedienerhandbuch
A-26
9650-1210-08 Rev. H
Tabelle 2. Genauigkeitstabelle der SCE-Niveaus und entsprechenden SPI-Schwellenwerte
Referenzliteratur:
[1] Eftestol T, Sunde K, Steen PA. Effects of Interrupting Precordial Compressions on the Calculated Probability of
Defibrillation Success during Out-of-Hospital Cardiac Arrest. Circulation 2002; 105:2270-2273.
[2] Sota Y, Weil MH, Sun S, Tang W. Xie J, Noc M, Bisera J. Adverse effects of interrupting precordial compression during
cardiopulmonary resuscitation. Critical Care Medicine 1997; 25:733-736.
[3] Young C, Bisera J, Gehman S, Snyder D, Tang W, Weil MH. Amplitude spectrum area: measuring the probability of
successful defibrillation as applied to human data. Critical Care Medicine 2004; 32:S356-S358.
SCE-
Niveau
SPI-Schwellen-
wert (mV-Hz)
Sensitiv-
ität%
Spezifizität%
HOCH
7.4
95
57
MITTEL
8.3
90
65
SCHWACH
10.8
80
80