WIKA CTD9300 Benutzerhandbuch

Seite 24

Advertising
background image

9037926 03/2005

Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

GB

24

WIKA Operating Instructions CTD 9300 Rev. 1

3.1.1.2 Display

The display menu is used for selecting the unit for displaying the
temperature and the desired calibration menu. In addition, the
averaging function of the temperature display can be set.

The desired menu item is
"enabled" using the cursor keys
and selected by pressing the
SELECT key. Selected units and
functions are identified by an
arrow.

Unit Selection
You can choose between degrees Celsius °C, degrees Fahrenheit
°F or Kelvin.

Calibration Menu Selections
If a measuring instrument is not integrated, you can choose
between the min./max. value and standard deviation menu items.
If a measuring instrument is integrated, the selection will be
supplemented with:
„

TC-Probe

„

Pt-Probe

„

mA-Probe

„

switch test

Only one menu can be selected at a time.

Min/max value
Apart from the current block temperature, the minimum and
maximum value of the block temperature will be displayed. Press
ENTER to set min. and max. to the topical measured value.
Using this function, it is easy to assess the maximum limits of
variation of the block temperature in stationary mode.

Fig. 11: Display adjustment