WIKA CTD9300 Benutzerhandbuch

Seite 26

Advertising
background image

9037926 03/2005

Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

GB

26

WIKA Operating Instructions CTD 9300 Rev. 1

3.1.1.3 System

This menu is used for setting the system functions.

Language Selection
In this menu, the user can
select one of the languages
German, English, French or
Spanish using the cursor keys.
The selection must be
confirmed with SELECT.

Time and Date
The system time and date are selected block by block using the
cursor keys and entered via the numeric keypad. [ hh -> mm -> ss ]
and [ 20yy -> mm -> dd ]. Confirm your entry with ENTER.
Display Settings
Brightness and contrast can be set via the numeric keypad within
the range of from 0 % to 100 %.
Confirm your entry with ENTER.
The following settings are recommended:
„

Brightness: 80 %

„

Contrast:

60 % (contrast settings < 50 % are not advisable)

Alarm
When the preselected alarm temperature is exceeded, "temp. to
high" flashes in the information line. The alarm temperature is
entered via the numeric keypad and must be confirmed with
ENTER.
Cut-out
When the preselected switch-off temperature has been exceeded,
the device will automatically be reset to the measuring mode, i.e.
the control unit is switched off. "temp. to high" will flash in the
information line. The device will cool to ambient temperature. The
switch-off temperature is entered via the numeric keypad and must
be confirmed with ENTER.

Fig. 12: System adjustment

Advertising