5.1

CTD 9300 from -30 to 165 °C

Minimum temperature

1)

Maximum temperature
Error of measurement up to 100 °C
Error of measurement above 100 °C
Stability up to -30 °C
Stability at 165 °C
Gradients axially, 4 cm, 165 °C
Gradients axially, 4 cm, -30 °C
Heating time from 20 °C to 165°C
Cooling time from 20 °C to -20 °C
Immersion depth
Sleeve diameter
Resolution in K
Footprint W x D
Footprint
Dimensions, W x D x H
Power consumption
Power supply

Fuse
Measurement value logging
Mass (without integ. measuring
instrument)
Ambient temperature
Standard ambient conditions

Overvoltages

Degree of pollution
Storage temperature
Humidity
Protection class
CE conformity

°C
°C
K
%
K
K
K
K
min
min
mm
mm
K
mm
cm²
mm
W
VAC

kg

°C

°C
% rel.h.

-30
165
0.1
0.1 from the measured value
< 0.05
0.01
< 0.06
< 0.04
12
7
160
28
0.01
160 x 320
512
160 x 320 x 420
400
100 ... 240 +10 %/-15 %
50/60 Hz

6.3A, slow blowing (T6.3A 250V)

optional
10

0...40
Use permissible indoors only
permissible altitude up to
2,000 m
Overvoltage category II
Transient overvoltage as it
normally occurs in the supply
network
Degree of pollution 2
-20 ... 40

0 ... 90
IP 20
EN61326-1, EN61010-1

1 )

1 )

1 )

1 )

1 )

at 23 °C ambient temperature

Advertising