WIKA CTD9300 Benutzerhandbuch

Seite 63

Advertising
background image

9037926 03/2005

63

WIKA Operating Instructions CTD 9300 Rev. 1

GB

Temperature Dry Well Calibrators CTD 9300
5. Technical Data

Cold junction
compensation

Resolution and
uncertainty
(Ambient temper-
ature T

U

= 25 °C)

Pt100 (DIN EN 60751), 4 wire technology:

-90.00 up to +850.00 °C: ±0.005 % of the final value
0.01 °C
NiCr/NiAl (DIN EN 60584-1):
-90.00 up to +999.99 °C: ±0.007 % of the final value
0.01 °C +1,000.0 to +1,370.0 °C: ±0.005 % of the
final value 0.1 °C
Pt10%Rh/Pt (DIN EN 60584-1):
0.00 up to +999.99 °C: ±0.05 % of the final value
0.01 °C +1,000.0 to +1,760.0 °C: ±0.03 % of the
final value 0.1 °C
Pt13%Rh/Pt (DIN EN 60584-1):
0.00 up to +999.99 °C: ±0.05 % of the final value
0.01 °C +1,000.0 to +1,760.0 °C: ±0.03 % of the
final value 0.1 °C
Pt30%Rh/Pt6%Rh (DIN EN 60584-1):
0.00 up to +999.99 °C: ±0.05 % of the final value
0.01 °C +1,000.0 to +1,820.0 °C: ±0.03 % of the
final value 0.1 °C
Fe/CuNi (DIN EN 60584-1):
-90.00 up to +900.00 °C: ±0.005 % of the final value
0.01 °C
Fe/CuNi (DIN 43710):
-90.00 up to +900.00 °C: ±0.005 % of the final value
0.01 °C
Cu/CuNi (DIN EN 60584-1):
-90.00 up to +400.00 °C: ±0.01 % of the final value
0.01 °C
Cu/CuNi (DIN 43710):
-90.00 up to +600.00 °C: ±0.01 % of the final value
±0.01°C
NiCr/CuNi (DIN EN 60584-1):
-90.00 up to +700.00 °C: ±0.005 % of the final value
±0.01 °C
Normal signal input:
±0.015 % of the final value ±0.01 mA
internal
for ambient temperatures between 0 and 60 °C

Advertising