Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

49

WIKA Operating Instructions CTD 9300 Rev. 1

3.3.2.3 Calibration Menu “Pt-Probe”

In this menu, the output signal
of a Pt resistance thermometer
can be displayed in addition to
the nominal and actual
temperatures of the calibrator.

Fig. 36: Calibration menu

“Pt-probe”

The following information is provided:

Designation
Probe

t-Probe

Error

Signal

Information
Type of probe

Temperature of the
probe
Error of the probe
in K
Electric output signal
of the probe

Note
Pt resistance thermometer:
Pt 100-3, i.e.
Pt 100 in 3-wire circuits
in this example: 102.99 °C

Deviation from the int. (ext.)
reference thermometer
In the case of the Pt resistance
thermometer, the resistance in
Ohm