WIKA CTD9300 Benutzerhandbuch

Seite 48

Advertising
background image

9037926 03/2005

Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

GB

48

WIKA Operating Instructions CTD 9300 Rev. 1

Designation
Probe
t-open

t-close

Hysteresis

Information
Type of probe
opening temperature
of the switch
closing temperature
of the switch
The difference between
"t-open" and "t-close"

Note
Temperature switch
in this example: 107.32 °C

in this example: 103.98 °C

in this case: 3.34 K

The following information is provided:

3.3.2.2 Calibration Menu “TC-Probe”

In this menu, the output signal
of a thermocouple can be
displayed in addition to the
nominal and actual
temperatures of the calibrator.

Fig. 35: Calibration menu

“TC-probe”

The following information is provided:

Designation
Probe

t-Probe

Error

Signal

Information
Type of probe

Temperature of the
probe
Error of the probe
in K
Electric output signal
of the probe

Note
Thermocouple: TCJ-3, i.e.
Thermocouple J at input 3
in this example: 102.67 °C

Deviation from the int. (ext.)
reference thermometer
In the case of the thermocouple,
the thermal voltage in µV