WIKA CTD9300 Benutzerhandbuch
Seite 62

9037926 03/2005
62
WIKA Operating Instructions CTD 9300 Rev. 1
Temperature Dry Well Calibrators CTD 9300
5. Technical Data
GB
5.3
Integratable Measuring Instrument CTI 9350
Resistance thermometer (RTD)
Pt100, Pt500, Pt1000 2, 3 or 4 wire technology
according to DIN EN 60751
Thermocouples (TC)
Fe/CuNi, type J according to DIN EN 60584-1
Fe/CuNiCr/NiAl, type K according to DIN EN
60584-1
NiCr/Ni, type N according to DIN 60584-1
Pt10%Rh/Pt, type S according to DIN EN 60584-1
Pt13%Rh-Pt, type R according to DIN EN 60584
Pt30%Rh-Pt6%Rh, type B according to DIN EN
60584
NiCr/CuNi, type E according to DIN EN 60584-1
Fe/CuNi, type L according to DIN 43710
Cu/CuNi, type T according to DIN EN 60584-1 Cu/
CuNi, type U according to DIN 43710
Normal signal input (mA)
0(4)-20 mA, programmable (the 24V DC transducer
supply can be enabled/disabled, I
max
= 30 mA)
Sensor inputs