WIKA CTD9300 Benutzerhandbuch

Seite 38

Advertising
background image

9037926 03/2005

Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

GB

38

WIKA Operating Instructions CTD 9300 Rev. 1

This menu comprises the parameters for linearising the internal
reference thermometer.

The internal reference
thermometer is linearised in two
stages. First of all, using a basic
linearisation that is always
performed and thereafter using
an individual fine linearisation.
In this case, "t'" is the
temperature following the basic
linearisation.

t

display = P0 + P1 * t' + P2 * t' ² + P3 * t' ³ + P4 * t' 4

Linearisation

CTD 9300-650

CTD 9300-165

parameters

P0

0

0

P1

1

1

P2

0

0

P3

0

0

P4

0

0

Using these parameters, only the general basic linearisation is
enabled and fine linearisation is switched off. In the case of DKD
calibration of the temperature calibrators, the errors of
measurement of the reference are exactly determined and the
parameters P0 to P4 are set to the optimum.

Fine linearisation can be disabled by setting the switch to OFF.
This function simplifies the determination of new parameters Pn
for adjusting the reference, at an autuorised laboatory.
The user should not edit these settings.

3.1.3.1 Internal Reference

Fig. 23: Internal reference

parameter set

Advertising