Temperature Dry Well Calibrators CTD 9300
4. Operating the Calibrator

GB

Corrective action

Check the applied mains
voltage against the
supply voltage specified
on the type label.
Pull the power cable,
check the fuse and
replace it if defective.
Check the ground wire
connection of the
calibrator.
Re-establish the
connection to the
ground wire system

Cause

External influences, e.g.
magnetic fields,
incorrect supply voltage

Ground wire monitor
active

Fault

Calibrator does not
work properly

Calibrator is working but
heater dos not work

Fault - What Can You Do?

If the device is still not working properly after you have checked the
above items, it must be checked at the manufacturer's.

4.6

Troubleshooting

Replacing the Fuses
The calibrators are equipped with a safety fuse. They are located on
the bottom of the calibrator, integrated in the power connection. If
the fan does not start of the display remains dark, check the fuse
and replace it, if it has blown. Replace defective fuses only with
fuses with the ratings specified in the present manual.

CTD 9300-165

CTD 9360-650

CTD 9360-650

100 ... 230 VAC

230 VAC

115 VAC

T6.3A 250V

T6.3A 250V

T10A 250V

If a replaced fuse blows repeatedly, the device is probably defective.
In this case, send the device to the manufacturer for repair.
In view of any warranty claims and your own safety, do not perform
repairs yourself.

Never open the housing!
The housing does not accommodate any components that can
be serviced by the user.