Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions
GB
31
WIKA Operating Instructions CTD 9300 Rev. 1
Date and Start time document the beginning of the recording.
Probe specifies the type of thermometer to be calibrated.
TC-Probe
Pt-Probe
mA-Probe
Menu item CL (Clear Logger) is
used to clear the stored
contents of the enabled set. If a
new calibration is started
without deleting the old data,
these data will be overwritten
with the new data.
Procedure when using the data logging function:
Select
ramp program in the Setup menu:
Select UP or DOWN with the cursor keys.
Change the desired temperature ramps
Set the tolerance to the desired value, e.g. 0.05 K
Set the duration to the desired value, e.g. 2 min
Set the cycles to 1
Select BACK to exit the
ramp program:
Select
Data Logger in the Setup menu:
Select the desired Set(n), n= 1, 2, ..., 8 with the cursor keys
If required, delete any calibration data stored in the respective
set with CL
Select ON with the cursor keys and enable the logging function
Select BACK to exit the
Data Logger
Press BACK again to return to the main menu (Calibrate).
Fig. 15: Recorded calibration
data