Temperature Dry Well Calibrators CTD 9300
3. Moduses and Functions

GB

31

WIKA Operating Instructions CTD 9300 Rev. 1

Date and Start time document the beginning of the recording.
Probe specifies the type of thermometer to be calibrated.
„

TC-Probe

„

Pt-Probe

„

mA-Probe

Menu item CL (Clear Logger) is
used to clear the stored
contents of the enabled set. If a
new calibration is started
without deleting the old data,
these data will be overwritten
with the new data.

Procedure when using the data logging function:
Select

ramp program in the Setup menu:

„

Select UP or DOWN with the cursor keys.

„

Change the desired temperature ramps

„

Set the tolerance to the desired value, e.g. 0.05 K

„

Set the duration to the desired value, e.g. 2 min

„

Set the cycles to 1

Select BACK to exit the

ramp program:

Select

Data Logger in the Setup menu:

„

Select the desired Set(n), n= 1, 2, ..., 8 with the cursor keys

„

If required, delete any calibration data stored in the respective
set with CL

„

Select ON with the cursor keys and enable the logging function

Select BACK to exit the

Data Logger

Press BACK again to return to the main menu (Calibrate).

Fig. 15: Recorded calibration

data