6.3.6
Status of measurement
Read status of measurement (object 6150h):
ID
DLC
Byte1
Byte2
Byte3
Byte4
Byte5
Byte6
Byte7
Byte8
600h + ID
8
40h
50h
61h
01h
0
0
0
0
Response from CANopen strain link:
ID
DLC
Byte1
Byte2
Byte3
Byte4
Byte5
Byte6
Byte7
Byte8
580h + ID
8
4Fh
50h
61h
01h
00h
0
0
0
Byte 5:
00h
actual measurement
03h
overflow of AD – converter
05h
underflow of AD – converter
The object 6150h is read only.
6.3.7
Process value 16bit
Read process value 16bit (object 7130h):
ID
DLC
Byte1
Byte2
Byte3
Byte4
Byte5
Byte6
Byte7
Byte8
600h + ID
8
40h
30h
71h
01h
0
0
0
0
Response from CANopen strain link:
ID
DLC
Byte1
Byte2
Byte3
Byte4
Byte5
Byte6
Byte7
Byte8
580h + ID
8
4Bh
30h
71h
01h
2Ch
01h
0
0
Byte 5 + 6:
2C’01h
01’2C (LSB first)
300
300 / 10
Number decimal digits
(6132) + unit(6131) = 3.00 µ
ε
The process value is quoted as 16-bit Integer double complement.
The object 2112h is changing the process value from 16bit (object 7130h) to 24bit (object 8130h) and vice
versa.
Example of negative strain:
ID
DLC
Byte1
Byte2
Byte3
Byte4
Byte5
Byte6
Byte7
Byte8
580h + ID
8
4Bh
30h
71h
01h
D4h
FEh
0
0
Byte 5 + 6:
D4’FEh
FE’D4 (LSB first)
-300
-300 / 10
Number decimal digits
(6132) + unit(6131) = -3.00 µ
ε