Manual_CANopen_StrainLink_EN_V3_05.doc

36/65

Baumer

22.07.11/dam

Frauenfeld, Switzerland

6.3.6

Status of measurement

Read status of measurement (object 6150h):

ID

DLC

Byte1

Byte2

Byte3

Byte4

Byte5

Byte6

Byte7

Byte8

600h + ID

8

40h

50h

61h

01h

0

0

0

0


Response from CANopen strain link:

ID

DLC

Byte1

Byte2

Byte3

Byte4

Byte5

Byte6

Byte7

Byte8

580h + ID

8

4Fh

50h

61h

01h

00h

0

0

0


Byte 5:

00h

actual measurement

03h

overflow of AD – converter

05h

underflow of AD – converter


The object 6150h is read only.

6.3.7

Process value 16bit

Read process value 16bit (object 7130h):

ID

DLC

Byte1

Byte2

Byte3

Byte4

Byte5

Byte6

Byte7

Byte8

600h + ID

8

40h

30h

71h

01h

0

0

0

0


Response from CANopen strain link:

ID

DLC

Byte1

Byte2

Byte3

Byte4

Byte5

Byte6

Byte7

Byte8

580h + ID

8

4Bh

30h

71h

01h

2Ch

01h

0

0


Byte 5 + 6:

2C’01h

01’2C (LSB first)

300

300 / 10

Number decimal digits

(6132) + unit(6131) = 3.00 µ

ε


The process value is quoted as 16-bit Integer double complement.
The object 2112h is changing the process value from 16bit (object 7130h) to 24bit (object 8130h) and vice
versa.

Example of negative strain:

ID

DLC

Byte1

Byte2

Byte3

Byte4

Byte5

Byte6

Byte7

Byte8

580h + ID

8

4Bh

30h

71h

01h

D4h

FEh

0

0


Byte 5 + 6:

D4’FEh

FE’D4 (LSB first)

-300

-300 / 10

Number decimal digits

(6132) + unit(6131) = -3.00 µ

ε


The object 7130h is read only.