Heater/gas – Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 105
Advertising
Manual control
79
Stop Pump 2.
Heater/Gas
The heating, gas flow pump and inert gas valve of the 774 Oven Sample Proces-
sor are switched on and off here.
Heater
Heats the oven to the required temperature for the defined heating duration.
Temperature
[ 50 ] ... 250 °C, Init
Temperature to which the oven is to be heated. With
Init
heating will be car-
ried out up to the initial temperature set on the device.
Heating period
[ 1 ] ... 999 min
Duration of the heating-up phase until the required temperature has been
achieved.
Start heating. The current temperature is shown in the status display.
Stop heating.
Advertising