Meas conc – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 511

Advertising
background image

Commands

485

Minimum evaluation Temperature

on, [ off ]

For the minimum measured value the associated time and temperature are

interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 °C/s

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the threshold value.

Maximum evaluation Temperature

on, [ off ]

For the maximum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 °C/s

The evaluation of the maximum begins as soon as the slope of the curve

exceeds the threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values


MEAS Conc

Command for concentration measurements (direct measurement).

Devices

This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

MEAS Conc

are configured on the following 3

tabs:

General/Hardware

Parameters for device, sensor and stirrer.

Measuring parameters

Parameters for the measurement.

Additional measured values

Defining further measured values which can be saved as additional col-
umns in the list of measuring points.

Advertising