410

Method

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of

the command.

Start conditions

Screenshot and parameters: see SET pH - Start conditions

Control parameters

Command name

25 characters, [ KFT Ipol # ]

Name of the command.

Endpoint

EP at

-2000.0 ... [ 250.0 ] ... 2000.0 mV, off

Measured value for the endpoint.

Titration rate

Titration rate

Three predefined sets of parameters can be selected for the titration rate:

slow

,

optimal

and

fast

. For these sets the parameters for the

Control

are not

displayed. In order to edit these parameters the value

user

must be se-

lected.

slow

For titrations in which the finest details are also to be visible. However, this

can also increase noise and lead to the production of unwanted EPs.

[ optimal ]

Parameter set for all standard titrations; optimized for the most frequent
applications.

fast

For less critical rapid titrations.

user

Editing the individual titration parameters which affect the titration rate.