Method

General/Hardware

The general parameters for the control device, the electrodes and the stirrer are

defined on this tab.

Command name

25 characters, [ KFC # ]

Name of the command.

Device

Device name

Device name, [ not defined ]

Selection of the device from device table. Only those devices being able to
execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ 831 KF Coulometer ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Indicator electrode

I(pol)

2, 5, [ 10 ], 20, 30 µA

Polarization current at the indicator electrode.

Note

On changing this parameter take care that the polarization current of the indica-
tor electrode, the endpoint and the control range are linked to each other (de-
tails see

Control parameters

).

Electrode test

on, [ off ]

Switching on/off the electrode test for polarized electrodes. The electrode

test for polarized electrodes is carried out during the transition from an in-
active condition to a measuring condition