Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 370

Advertising
background image

344

Method

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Dosing device

Dosing device

[ 1 ] ... 4

(808, 809, 835, 836, 857)

[ 1 ] ... 3

(855)

[ internal D0 ]

(Titrino)

external D1, external D2

(only 736, 751, 758, 799)

Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing.

Solution

24 characters, solution names, [ not defined ]

Input or selection of the titrant or solution defined in the solution table. Addi-
tionally,

not defined

can be selected. If intelligent exchange or dosing units

are used then a check will be made in the method sequence that the cor-
rect titrant is present in the connected dosing device and whether the type
of dosing device is correct. With non-intelligent exchange or dosing units
the cylinder volume is checked. At the start of the command the validity of
the titer and the expiry date of the selected titrant will be checked as well as

the GLP test interval for the buret. If

not defined

is selected no tests will be

carried out.

Sensor

Measuring input

[ 1 ] ... 2

(808, 809, 835, 836, 857)

[ 1 ]

(855)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ Metal electrode ]

Selection of a sensor of type

pH electrode, Metal electrode

or

ISE electrode

from the list of sensors defined in the sensor table. The calibration data for

the ISE and pH sensors will be adopted for the determination.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Titration pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.