4.11.3 Monitoring

report

If the method contains commands of type

STAT

or

DOS

, with the context-sensitive

menu item

Monitoring report...

in the curve window the dialog window

Monitoring

report - 'Command name'

opens for showing information about limit violations for

each monitoring switched-on.

The following columns are displayed:

Time [s]

Time in s until limit violation.

Limit

lower

Lower limit exceeded.

upper

Upper limit exceeded.

ok

Measured value is inside limits again.

[pH] / [U]

Measured value limit.

Meas.v. [pH] / Meas.v. [U]

Measured value at limit violation time.

Volume [mL]

Dosing volume at limit violation time.

Temp. [°C]

Temperature at limit violation time.

Action

The running

STAT

command was quit, then the determination was stopped.

Stop command

(only Titrando)

The running

STAT

command was quit, then the next command was started.

Wait for [Continue]

Reagent dosing in the current

STAT

command was interrupted until it was

continued by a manual command.

Wait for limit ok

Reagent dosing in the current

STAT

command was interrupted until the

monitored measured value was again within the limits (including hystere-
sis). Afterwards reagent dosing was resumed automatically.

[ none ]

No action was taken.