Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 728

Advertising
background image

702

Configuration

Edit rack properties (774)

Rack name

Shows the name of the attached rack. If no rack is in position then "

-----

" will

be shown.

Rack code

Shows the rack code of the attached rack. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the

Sample Processor in order to recognize the rack. If no rack is in position
then the display will be empty.

Number of positions

Shows the number of positions on the attached rack. If no rack is in posi-
tion then the display will be empty.

The parameters for the attached rack are defined on the following 2 tabs:

Lift positions

Defines the work, rinse, shift and special positions for Tower 1.

Special beaker

Settings for the special beaker on the rack in position.

Lift positions

Tower 1

Defines the lift positions for Tower 1. These apply for all rack positions ex-
cept those that are defined as Special beaker.

Work position

0 ... [ 83 ] ... 100 mm

Work position for Lift 1. At this lift position the electrodes, stirrer and buret
tips are optimally positioned for work.

Rinse position

[ 0 ] ... 100 mm

Rinse position for Lift 1. This lift position is used for rinsing the electrodes.