Subwindow Devices

681

Swing direction

+

(clockwise rotation), – (counterclockwise rotation)

The swing direction of the robotic arm depends on the type of robotic arm.
For a 2-tower model the robotic arm at Tower 1 must swing to the right, i.e.
"–" and that at Tower 2 must be mounted "

+

" so that it swings to the left.

External position

In the dialog window

External position #

a swing angle and a specific work position

can be defined for each external position. Shift and rinse positions can only be de-
fined for all 4 external positions together (see Tower # properties). No special po-
sition can be defined for external positions.

Angle

Offset...[ 60.00 °]...(Offset + max. swing range)

Definition of the turning angle for the selected external position. The offset
is made up of a design-dependent angle (approx. 8-9°) together with the
robotic arm offset from the Robotic arm properties (see Robotic arm prop-
erties)
. The maximum swing range is also defined under Robotic arm prop-
erties.

Work position

[ 0 ] ... 235 mm

Definition of the work position for the selected external position.

Rack

On the tab

Rack

the rack-specific data of the attached rack are shown.

Rack name

Shows the name of the attached rack. If no rack is in position then "

-----

" will

be shown.

Rack code

Shows the rack code of the attached rack. The rack code corresponds to
the arrangement of magnets on the base of the rack and is read in by the