Reprocessing rules – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 267

Advertising
background image

Subwindow Determination overview

241

lation a progress bar appears; the recalculation can be canceled with

alongside the bar. The results of this recalculation are entered automatically
in the subwindow

Result view

. This button is inactive until modifications

have actually been made. After the recalculation further data can be modi-
fied and then recalculation triggered again.

With this button all the modifications that have been made in reprocessing
and that have not been saved can be rejected; the original data and results

are then available again. This button is inactive until modifications have ac-
tually been made or when the recalculation has not yet been triggered.

With this button each determination that has been modified by reprocess-
ing will be saved as a new version with a version number increased by

+1

and the subwindow

Reprocessing

will be closed. This button is inactive when

the recalculation has not yet been triggered and if not all the selected de-
terminations could be reprocessed.

With this button the result of the reprocessing carried out since the last data

saving will not be saved and the subwindow

Reprocessing

will be closed.

The same function can also be triggered with in the right top corner of
the window.

Reprocessing rules

The following rules apply for reprocessing the selected and modified determina-
tions:

Number of determinations

No more than 100 determinations (including statistically linked determina-

tions) may be selected for reprocessing.

Statistically linked determinations

If determinations are selected for reprocessing that are statistically linked
with other determinations that have not been selected then these determi-
nations will also be automatically reprocessed and then saved as new ver-
sions. Determinations that are statistically linked with each other must al-

ways be reprocessed in the same sequence as they were recorded. If no
modification is made to a determination then this determination will retain
its original determination status (i.e.

original

for non-modified determina-

tions).

Start test

During the start test all device tests and device monitoring will be ignored.

Special tracks

During reprocessing the series start, series end and error tracks are not run
through again.

Tracks with "Return immediately"

If the method contains tracks for which the option

Return immediately

has

been switched on then a warning will appear to say that it can no longer be
guaranteed that the method will proceed in the same way during reproc-
essing as during the determination itself.

Advertising