Commands

443

Control parameters

Command name

25 characters, [ STAT U # ]

Name of the command.

Control point

Control point at

-2000.0 ... 2000.0 mV, [ aus ]

(Titrando),

-2000 ... 2000 mV, [ aus ]

(Titrino)

Definition of the current value to be controlled.

Titration rate

Titration rate

50 µL/min, [ 100 µL/min ], 500 µL/min, user

Three predefined sets of parameters can be selected for the titration rate:

50 µL/min

,

100 µL/min

and

500 µL/min

. For these sets the parameters for

the

Control

are not displayed. In order to edit these parameters the value

user

must be selected.

Control

(only visible for

Titration rate

=

user

)

Dynamics

0.1 ... [ 60.0 ] ... 2000.0 mV, off

(Titrando)

1 ... [ 60 ] ... 2000 mV, off

(Titrino)

Dynamics defines the measured value range before the given endpoint.
This parameter has an important influence on the titration rate and with it on
the accuracy. Within the control range the steps which are added are con-

trolled by the

Min. rate

. The nearer the endpoint the slower the dosing takes

place until the

Min. rate

has been reached. Outside the range the dosing

takes place with

Max. rate

) (see Control range).

Max. rate

0.01 ... [ 0.25 (50 µL/min) ] ... [ 0.75 (100 µL/min) ] ... [ 2.00 (500 µL/min) ] ...
166.00 mL/min, maximum

(Titrando)

0.01 ... [ 0.25 (50 µL/min) ] ... [ 0.75 (100 µL/min) ] ... [ 2.00 (500 µL/min) ] ...
150.00 mL/min, maximum

(Titrino)

Speed of addition outside the control range. The maximum dosing speed
depends on the cylinder volume of the exchange unit or dosing unit used.
Rule of thumb for the maximum rate:

Max. rate

in mL/min = 0.005 • Ex-

pected rate of reaction in µL/min