Commands

421

Start drift

1 ... [ 20 ] ... 999 µg/min

The determination can only be started when the measured drift is below the

entered value. The Status in the

Run

window (Single determination or De-

termination series) then switches to

COND READY

.

Drift correction

automatic, manual, [ off ]

If this parameter is set to

automatic

or

manual

the drift correction ( =

Drift

value

x Titration time) caused by blank consumption during titration is de-

ducted from the total consumption. With

automatic

the drift value at the start

of the titration applies with

manual

the value defined under

Drift value

is ap-

plied.

Drift value

[ 0 ] ... 99.9 µL/min

Drift value for drift correction. This parameter is only shown for

Drift correc-

tion

=

manual

.

Delay after 'Cond ok'

[ 0 ] ... 999 s

Waiting period until the start of the titration during which the conditioning
condition must be fulfilled.

Sample addition time

[ 0 ] ... 99 s

Waiting time between stop of conditioning and start of titration during which

the sample can be added.

Only start titration by a start command from a SEND command

on, [ off ]

If this option is set the titration is only started when the

Event message

Start

titration

from a

SEND

command reaches this command.

Note

At the time of reception of the

SEND

command conditioning must already be

terminated (=

Condok

) otherwise the command to start the titration will be igno-

red.

Show measured value during conditioning

on, [ off ]

If this option is set additionally to the drift and time the measured value is
shown in the Live display.