Subwindow Sensors

747

the graph in orange for warning limits and red for intervention limits. However,
whether these limits are observed is not monitored, i.e. infringing these limits does

not trigger any action. For monitoring during calibration the limits set under Limits
(see Sensors - Properties - Limits) apply.

Warning limits for slope

on, [ off ]

If this option is switched on then the values for the slope will be shown in
orange if the limits are infringed.

Lower limits

-2.147E+9 ... 2.147E+9

pH electrode:

97.5 %

ISE (+1):

57.2 mV

ISE (-1):

-61.2 mV

ISE (+2):

27.7 mV

ISE (-2):

-31.7 mV

ISE (+3):

17.7 mV

ISE (-3):

-21.7 mV

ISE (+4):

12.8 mV

ISE (-4):

-16.8 mV

Lower warning limits for the slope.

Upper limits

-2.147E+9 ... 2.147E+9

pH electrode:

102.5 %

ISE (+1):

61.2 mV

ISE (-1):

-57.2 mV

ISE (+2):

31.7 mV

ISE (-2):

-27.7 mV

ISE (+3):

21.7 mV

ISE (-3):

-17.7 mV

ISE (+4):

16.8 mV

ISE (-4):

-12.8 mV

Upper warning limits for the slope.

Intervention limits for the slope

on, [ off ]

If this option is switched on then the values for the slope will be shown in

red if the limits are infringed.

Lower limits

-2.147E+9 ... 2.147E+9

pH:

95.0 %

ISE (+1):

55.0 mV

ISE (-1):

-62.4 mV

ISE (+2):

26.7 mV

ISE (-2):

-32.7 mV

ISE (+3):

16.7 mV

ISE (-3):

-22.7 mV

ISE (+4):

11.8 mV

ISE (-4):

-17.8 mV

Lower intervention limits for the slope.

Upper limits

-2.147E+9 ... 2.147E+9

pH:

105.0 %

ISE (+1):

62.4 mV

ISE (-1):

-55.0 mV

ISE (+2):

32.7 mV

ISE (-2):

-26.7 mV

ISE (+3):

22.7 mV

ISE (-3):

-16.7 mV

ISE (+4):

17.8 mV

ISE (-4):

-11.8 mV

Upper intervention limits for the slope.