Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 520

Advertising
background image

494

Method

continuous

A temperature sensor must be connected. The temperature is measured

continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(see

Meas-

uring parameters

) will be used.

off

The temperature will not be measured. The

Temperature

(see

Measuring pa-

rameters

) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

Setting the stirrer speed. The direction in which stirring takes place changes

as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

The stirrer is automatically switched off at the end of the command if this
option is enabled.

Standard addition

The standard addition parameters are defined on this tab.

Command name

25 characters, [ STDADD man # ]

Name of the command.

Number of additions

1 ... [ 3 ] ... 19

Number of times the standard solution is to be added.

Volume auxiliary solution

[ 0.000 ] ... 9999.999 mL

Volume of the auxiliary solution (e.g. buffer) to be added to the sample be-
fore the first standard addition.