Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 720

Advertising
background image

694

Configuration

Opens the dialog window

Load new program version

.

Old version

Shows the old device program version.

New version

Shows the new device program version to be loaded.

Loads new program version.

Note

Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.

Towers

On the tab

Towers

the tower parameters for Tower 1 and Tower 2 (if present) are

shown for the 730 Sample Changer. These cannot be edited, but must be set on
the device itself via the keyboard.

Tower 1 / Tower 2

Max. stroke path

Shows the lowest permitted lift position for the two towers. A lift height of

0

mm

corresponds to the "Rest position", i.e. the lift is moved right to the top.

Beaker sensor

Shows whether the beaker sensor for the two towers is switched on or off.
Each time that a sample position is moved to the beaker sensor checks

whether a beaker is present or not.

Tower 1

Number of pumps

Shows how many pumps are connected to Tower 1.

759 Swing Head

Shows whether a swing head is mounted on Tower 1.

Tower 2

Number of pumps

Shows how many pumps are connected to Tower 1.

Note
A swing head can only be mounted on Tower 1.

Advertising